A quadrupole mass analyzer works by combining a radio frequency (RF) alternating current (AC) potential with a direct current (DC) potential over four electrodes, or poles, to create the electric field that sample ions pass through. As the ions pass through this electric field, they gain energy and accelerate. Two of the oppositely placed poles have a positive potential applied to them, and the other two have a negative potential applied. So, the quadrupole actually generates two overlapping electric fields: one in the horizontal plane and the other in the vertical plane.
The DC component of the electric field produces a constant force on the ions while they travel through the quadrupole. Because the ions are positively charged from the plasma, the positive poles repel the ions into the center of the mass analyzer, while the negative poles attract the ions away from the center of the quadrupole. The AC component counteracts the effect of the DC component depending on the ion’s mass. Because the ions gain energy as they pass through the quadrupole, the degree to which the AC component affects the ion flight path increases along the length of the quadrupole.
Lighter ions are more easily affected by the AC field than heavier ions and they follow the oscillations of the RF driving circuit. Heavier ions are significantly less responsive to the AC field, tending to follow the DC field instead. This means that for the positive pair of poles, lighter ions are pushed out of the quadrupole along its length by oscillations. For the negative pair of poles, the lighter ions are held in the quadrupole by oscillations
For both the vertical and horizontal electric fields, only specific combinations of AC and DC potential result in a stable flight path through the quadrupole for any given mass. Plotted on a stability diagram, there is overlap in the stability regions, so the ion can pass through the quadrupole mass filter.
The size of this stability region is different for each ion mass, so individual masses can be resolved with specific AC/DC potential combinations. Mass resolution is controlled by the ratio of DC against AC potential, and unit mass resolution is easily achieved. Keeping the ratio constant and increasing the potentials allows each mass in the spectrum to be scanned individually.
Mass resolution can be controlled by increasing the DC offset of this line, but this does come with an associated loss in sensitivity. Also, the practical resolution limit depends on the overall sensitivity of the instrument. Typically, this is 0.7–0.85u in ‘normal’ resolution. However, in most quadrupole mass spectrometers, resolution can be switched to ‘high’ resolution at 0.3–0.4u. Such a switch helps attenuate very high signals of, for example, matrix components.